Hi Scan Touch

Parameter Specifications

Model
Pulsar2 Minimal Stress

PROGRAMMABILITY

Programs
over 1000 user programs, each including 3 U/S, 2 DIATH, 2 I/A, 2 VIT modes

EQUIPMENT CONTROL

User Interface
Graphic User Interface, 12.1” LCD with touch screen
Remote control
cable wired, controls all functions
Footpedal
dual linear programmable footpedal controls all equipment functions
Automatic cart for bottle
programmable and controlled by pedal and console

IRRIGATION

Fluid delivery
gravity fed - eye pressure determined by the height of I.V. pole
patented forced irrigation - eye pressure determined by programmable air pressure (range 14-80 mmHg)
Valving element
solenoid driven pinch valve
Control
system footpedal/console

ASPIRATION

Aspiration pump types
Venturi and Peristaltic
Actuating medium
pressurized air from external source
pressure: from 500 to 800 KPa (72 to 116 PSI)
flow: at least 30 Normal litre/minute @500 mmHg
Default vacuum level
user programmable
Available vacuum range
5 to 500 mmHg
Default flow rate
user programmable
Flow rate range
2 to 50 cc/min (peristaltic only)
Aspiration rise time
three levels: Slow, Medium, Fast
Surgeon mode
(Linear Aspiration): Linear aspiration
(Vacuum and/or flow rate) from zero to preset
linearly controlled via system footpedal
Safety system
ACS3 (Anterior Chamber Surge Suppression System)
Control
system footpedal

VITRECTOMY

Handpiece type
20/23/25G pneumatically powered guillottine cutter (VIT)
Cutting mode
reciprocating motion
Default cut rate
user programmable
Cutting rate
from 60 up to 4000 cuts per minute (VIT)
Port size
adjustable from 0.2 to 0.7 mm (reusable cutters only)
Actuating medium
pressurized air from external source
Operating pressure
230 KPa (33,4 PSI)
Surgeon mode
linear cut rate from 0 to preset controlled via system footpedal
Control
system footpedal

DIATHERMY

Type
Bipolar generator. Generator stops when RF power is not needed
Operating frequency
2 MHz
Nominal power
7 W (450 Ohm)
No load max. voltage
100 V
Default bipolar power
user programmable
Bipolar power
5 to 100%
Surgeon mode
allows linear control of the diathermy power via system footpedal depression
Handpiece type
(Linear power): Bipolar microforceps, slim stat pencil eraser, intraocular diathermy pencils
Control
system footpedal

PHACOEMULSIFIER

Handpiece type
piezoelectric
Frequency
approx. 40 KHz
Tip stroke
from 0 to 100 μm
Power
panel or linear control of the U/S power via system footpedal depression
U/S Mode
linear or panel; Continuous, Short Pulse, Single Burst, Multi Burst, Continuous Burst, Programmable Emission Mode
U/S Timer
from 0.00 minutes to 9.59 minutes - equivalent Phaco time display
Control
system footpedal

ELECTRICAL SPECIFICATIONS

Dimensions
330 x 410 x 450 mm
Weight
15 Kg
Input voltage
100/120/220/230-240 V A.C. SELECT.
Frequency
50/60 Hz
Power consumption
145 W
Input air pressure
from 500 to 800 KPa (72 to 116 PSI)

REGULATORY

CE marking
european directive 93/42/EEC
Safety
conforms to IEC 60601-1
IEC 60601-1-2
IEC 60601-2-2