Keratron Nova

Parameter Specifications

Keratron™ Nova


Mountable on the most popular slit lamps
Area of analysis
10mm x 14 mm
Keratoscope cone
28 border mires equally spaced on a 43D sphere. Bivalent cone with Near/Far selection by touch screen
Analysed points
over 80,000
Measured points
Corneal coverage
From 0.33 mm (minimum diameter on a 43D sphere) up to 11mm on a normal eye
Measured area
90% of the corneal surface (normal eyes) with "close" cone selection. 75% of the corneal surface (normal eyes) with "distant" cone seletion
Dioptric power range
From 1 to over 120D
± 0,01D - 1 micron
Focusing device
Eye Positioning Control System EPCS (patented) automatic acquisition with decentration correction
Monocromatic - high sensitivity (CCD)
5.7'' TFT display with touchscreen
Other features
Processing on Videokeratoscope of rings, axial, curvatures, spherical offset and Gaussian maps, TBUT, sim-K, E-values, pupil data. Internal patients database
Accessories included
Calibration sphere, touch pen, USB memory stick, Software Scout, footswitch, power supply, Ethernet cross-over cable
Capturing images
Capturing by footswitch or button. Capturing in semi-automatic mode ("burst")
Communication external PC and VK
Via Ethernet cable (cross) provided for connecting "peer to peer" with the external PC (not supplied), or via a LAN network. Exams export to USB stick
Power Supply
100-240VAC, 50/60Hz, with medical power supply provided

Computer (recommended Minimal Requirements)

Operating system
MS Windows 7/8/10 (32 e 64 bit)
Intel I3, 2GB RAM
Internal 180 GB HD, CD-ROM
Super VGA color monitor 14", 1024x768 points 16 millions color
Color printer
Ethernet, USB 2.0 High speed


Dioptric scale
Absolute, Normalized, Adjustable
Keratometric values and indices
K-readings. Meridians, semi-meridians, Maloney indices, eccentricity, CLMI Keratoconus indices, keratoconus follow-up
Definition of the edge, diameter and decentering (angle K)
Zones and grids
Indication of the diameters 3, 5 and 7 mm, Cartesian axes, and millimeter grid
Local curvature, axial curvature, wavefront OPD or wavefront error (WFE) refraction and height map with 3D insert
Axis moving
Position of the axis selectable as corneal vertex, pupil center on any other choice
print screen with header of the institution, or personalized print templates
Special functions
Profiles, difference, repeatability check, maps comparison caliper, refraction calculator
Image acquisition and videos
Enabled with TV camera mounted on slit lamp and any video capture board (not included)
On-line help
Detailed on line multi language help detailed for all functions
Contact lens
Fluorescein pattern simulation of most international contact lens manufacturer geometries. Tilting to simulate lid pressure. Lens displacement in any position. Eccentricity measure at 6 and 8 mm, over refraction calculator. Personalised auto-fit for customised lens. Ortho-K custom fitting. Adjustable clearance scale. Link to third party software.
Internet connection
Maps can be sent as attachments to e-mail messages
Local network and database
Management of one or more independent databases shareable in a network


Analysing contact lenses kit Base with joystick and left/right sensor Standard mires cone


CE mark
Directive 93/42/CEE