Keratron Scout - Technical specifications
Configurations
PORTABLE:
Battery or cable operated
SLIT LAMP:
Easily mounted on any slit lamp
INTRA-OPERATIVE:
Balanced arm surgical trolley, disposable sterile plastic covers
Videokeratoscope
AREA OF ANALYSIS:
10mm x 14mm (visible on the monitor)
KERATOSCOPE CONE:
28 border mires, equally spaced on a 43D sphere
ANALYSED POINTS:
Over 80,000
MEASURED POINTS:
7,168
CORNEAL COVERAGE:
From 0.33mm (minimum diameter on a 43D sphere) up to 11mm on a normal eye
MEASURED AREA:
90% of the corneal surface (normal eyes)
DIOPTRIC POWER RANGE:
From 1D to over 120D
RESOLUTION:
+/- 0.01D, 1 micron
FOCUSING DEVICE:
Eye Positioning Control System EPCS (patented) automatic acquisition, with decentration correction
TV CAMERA:
High resolution (C.C.I.R.)
MONITOR:
4" B&W
OTHER FEATURES:
"OK" button and OD/OS acquisition keys, reverse OR keys, low-power standby function, tiltable mires cone (0°-10°)
WEIGHT:
1 Kg approx.
ACCESSORIES INCLUDED:
Calibration set, Scout software
Computer (recommended minimal features)
ENVIRONMENT:
MS Windows 98/Me o 2000/XP
PROCESSOR, MEMORY:
Pentium III 450MHz, 64 Mb RAM
DISKS:
Internal 10 Gb HD, internal 8x CD-Rom, Floppy disk drive 3 1/2'' - 1.44Mb
MONITOR:
Super VGA color monitor 14'', 1024x768 points, 16 million colors
PRINTER:
Ink jet (eg. HP 690-990 or equivalent)
PORT:
USB Port
Software
DIOPTRIC SCALE:
Absolute, Normalized, Adjustable
KERATOMETRIC VALUES AND INDICES:
K-readings, Meridians, Hemimeridians, Maloney Indices, Eccentricity
PUPIL:
Border detection, diameter and decentration
ZONES AND GRIDS:
3, 5 and 7 mm, orthogonal axis or millimeter grid
MAPS:
Local curvature, axial curvature, wavefront OPD
or Wfe refraction map with 3D insert
MOVE AXIS:
Position of the axis selectable as corneal vertex, pupil center or any other choice
PRINT:
Print screen with header of the institution, or personalized print templates
SPECIAL FUNCTIONS:
Profiles, difference, repeatability check, maps comparison, caliper, refraction calculator
IMAGE ACQUISITION AND MOVIES:
Enabled with TV camera mounted on a slit lamp and video capture board (not included)
ON-LINE HELP:
On-line multi-language help detailed for all functions
CONTACT LENS:
Fluorescein pattern simulation of most of international contact lens manufacturers' geometries.
Tilting to simulate lid pressure.
Lens displacement in any position.
Eccentricity measured at 6 and 8 mm,
over-refraction calculator.
Personalized auto-fit for customized lens.
Adjustable clearance scale.
Link to third party software.
INTERNET CONNECTION:
Maps can be sent as attachments to e-mail messages.
LOCAL NETWORK AND DATABASE:
Management of one or more independent databases shareable in a network.
Accessories
DOCKING BASE:
Power supply for videokeratoscope, USB interface to the computer, battery charger
SLIT LAMP KIT:
Slit lamp attachment for tonometer hole D=8mm, connection cable
OR TROLLEY:
Balanced arm trolley and PC interface
FAR MIRES CONE:
For deep set eyes. 28 mires are 3mm more recessed than standard cone.
OR DISPOSABLE PLASTIC COVERS:
Disposable sterile plastic covers for standard and/or far cone
MEMORY/BATTERY CHARGER EXPANSIONS:
Internal boards for hand held battery operated use to capture up to 8 pictures.
BATTERY CHARGER:
Normal (15h) or fast (2h).
BATTERY/HEAD
REST MODULE:
Can be connected to videokeratoscope for hand held operation. Battery 1200mAh (1 hour)
Regulatory
CE MARK:
DIRECTIVE 93/42/CEE